WTW laboratory conductometer inoLab® Cond 7110

The WTW laboratory conductometer inoLab® Cond 7110 is a high-precision measuring instrument specially developed for the determination of conductivity, salinity and TDS (Total Dissolved Solids) in various liquids. It is ideal for use in laboratories that require accurate and reliable measurements. Thanks to its user-friendly operation and compact design, it is an indispensable tool for professionals in water and environmental analysis, the food and beverage industry and in chemical and pharmaceutical research.

Precise and reliable measurements

The inoLab® Cond 7110 offers impressive measurement accuracy and stability. It covers a wide conductivity range from 0 µS/cm to 1000 mS/cm and offers an accuracy of ±10.5 % of the measured value. The salinity measurement ranges from 0.0 to 70.0 (according to IOT) and the TDS measurement from 1 to 1999 mg/l. The temperature measurement is also precise and ranges from -5.0 to 105.0 °C with an accuracy of ±0.1 °C.

Similar to the WTW conductivity pocket meter ProfiLine Cond 3110, the inoLab® Cond 7110 offers high measurement accuracy and reliability, making it an excellent choice for demanding applications.

Ease of use and flexibility

The device is extremely user-friendly and does not require extensive training of laboratory personnel. The operating instructions are clear and easy to understand so that all important functions can be learned quickly. If required, a team of experts is available for a quick start-up.

The inoLab® Cond 7110 has various cell constants (0.450…0.500 cm-¹, 0.09…0.110 cm-¹, 0.800 to 0.880 cm-¹, 0.25…2.5 cm-¹, fixed 0.01 cm-¹), which allow flexible adaptation to different measurement requirements. The 1-point calibration and temperature compensation (nLF, linear 0.000 to 3.000 %, can be switched off) also contribute to the high flexibility and accuracy of the device.

Certifications and standards

The WTW laboratory conductometer inoLab® Cond 7110 complies with all relevant standards and certifications required for use in laboratories. It is manufactured to the highest quality standards and guarantees reliable and reproducible results.

Application examples

The inoLab® Cond 7110 is used in a wide range of applications:

  • Water and environmental analysis: Determination of conductivity and salinity in water samples.
  • Food and beverage industry: Monitoring of water quality and control of production processes.
  • Chemical and pharmaceutical research: analyzing solutions and checking the purity of chemicals.
  • Education and research: Use in laboratories at universities and research institutes for education and research.

FAQs

How do I calibrate the inoLab® Cond 7110?

  • The device has a 1-point calibration. Follow the instructions in the user manual to perform the calibration.

Which cell constants are available?

  • The device offers various cell constants: 0.450…0.500 cm-¹, 0.09…0.110 cm-¹, 0.800 to 0.880 cm-¹, 0.25…2.5 cm-¹, fixed 0.01 cm-¹.

Can the device also measure TDS?

  • Yes, the inoLab® Cond 7110 can measure TDS in the range from 1 to 1999 mg/l.

Is the device suitable for use in the food industry?

  • Yes, the device is ideal for monitoring water quality and controlling production processes in the food and beverage industry.

Conclusion

The WTW laboratory conductometer inoLab® Cond 7110 is a versatile and precise measuring device that can be used in various industries. It offers high measurement accuracy, ease of use and flexibility, making it an excellent choice for laboratories that require reliable and accurate measurements. Comparisons such as the WTW conductivity pocket meter ProfiLine Cond 3310 show that the inoLab® Cond 7110 offers excellent performance and reliability.

Analytical device: conductivity meter
Shipping weight: 2,32 kg
Item weight: 2,32 kg
WTW Laboratory Conductivity Meter inoLab® Cond 7110
Conductivity 0 µS/cm…1000 mS/cm ±10.5 % of value
Salinity 0.0…70.0 (acc. to IOT)
0.00…20 MOhm cm
TDS 1…1999 mg/l
Temperature -5.0…105.0 °C ±0.1 °C (23…221 °F)
Cell constant 0.450…0.500 cm-1,
0.09…0.110 cm cm-1,
0.800 bis 0.880 cm-1,
0,25…2,5 cm-1, fix 0,01 cm-1
Calibration 1-point
Tref 20 °C/25 °C (68 °F/77 °F)
Temperature compensation nLF, linear 0.000 to 3.000 %, can be switched-off

all values ±1 digit