BINDER KB 53 cooling incubator

The BINDER KB 53 cooling incubator is a highly precise and reliable unit that has been specially developed for use in laboratories, medical facilities and scientific research centers. With a generous interior volume of 53 liters and precise temperature control, this incubator offers a wide range of possible applications. Thanks to the apt.line preheating chamber technology and a temperature range of -5 to 100 °C, the KB 53 is ideal for the storage and incubation of temperature-sensitive samples. Similar to the BINDER KT053 cooling incubator, the KB 53 offers a robust design and high reliability.

Technical data and functions

The BINDER KB 53 cooling incubator is characterized by its robust design and precise temperature control. Here are some of the most important technical data and functions:

  • Temperature range:-5 to 100 °C
  • Interior volume:53 liters
  • Maximum load:50 kg
  • Spatial temperature deviation at 37 °C:±0.2 K
  • Temporal temperature deviation at 37 °C: ±0.1 K
  • Heating time to 37 °C: 5 minutes
  • Energy consumption at37 °C: 70 Wh/h
  • Sound pressure level:49 dB(A)
  • Dimensions (W x H x D):635 x 835 x 580 mm (outside), 400 x 400 x 330 mm (inside)

The incubator has two slide-in grids, each of which can carry a maximum load of 15 kg. The door is fitted with a high-quality seal, which ensures a high level of insulation and thus supports temperature stability inside the appliance. The powerful cooling system ensures rapid cooling of the samples, which is particularly advantageous in time-critical applications.

Operation and control

Operation of the BINDER KB 53 cooling incubator is simple and intuitive. The incubator is equipped with a modern controller that offers various programs and functions. All settings can be made conveniently via the clear display, including real-time function and adjustable ramp function. An RS 422 interface enables uncomplicated data transfer to printers or other devices so that evaluations can be saved or printed directly.

Certifications and standards

The BINDER KB 53 cooling incubator meets the requirements of DIN 12880:2007 and complies with BINDER’s high quality standards. The temperature data was determined in accordance with the BINDER factory standard and is based on the recommended wall distances of 10% of the height, width and depth of the interior. These certifications and standards guarantee the reliability and precision of the unit.

Application examples

The BINDER KB 53 cooling incubator is versatile and is ideal for various applications in laboratories and research facilities. Here are some specific application examples:

  • Microbiological investigations:The incubator provides ideal conditions for cultivating microorganisms at constant temperatures.
  • Storage of temperature-sensitive samples:Samples such as enzymes, antibodies and cell cultures can be stored safely and stably.
  • Pharmaceutical research:The KB 53 is suitable for stability tests and the storage of pharmaceutical products under controlled conditions.
  • Food analysis:The incubator can be used to test food for microbiological contamination.

FAQs

How precise is the temperature control of the BINDER KB 53 cooling incubator?

The temperature control of the BINDER KB 53 cooling incubator is extremely precise. The spatial temperature deviation at 37 °C is only ±0.2 K, and the temporal temperature deviation is ±0.1 K.

How long does it take for the incubator to heat up to 37 °C?

The heating time to 37 °C is only 5 minutes, so you can start your experiments quickly.

What maximum load can the incubator support?

The BINDER KB 53 cooling incubator can support a maximum total load of 50 kg. The maximum load per shelf is 15 kg.

Is the incubator energy-efficient?

Yes, the BINDER KB 53 cooling incubator is particularly energy-efficient. The energy consumption at 37 °C is only 70 Wh/h.

Conclusion

The BINDER KB 53 cooling incubator is a powerful and reliable unit that is characterized by its precise temperature control, robust design and simple operation. With a generous interior volume and a wide range of possible applications, it is the ideal solution for laboratories and research facilities. Similar to the Memmert Peltier-cooled incubator IPPeco, the KB 53 offers high energy efficiency and reliability.

Load max.: 11 – 50 kg
Width outside (incubators,…): 635 mm
Width inside (incubators,…): 400 mm
Maximum temperature (heat): + 100 °C
Height outside (incubators,…): 835 mm
Height inside (incubators,…): 400 mm
Incubator model: cooling incubator
Interior volume: 51 – 100 l
Depth outside (incubators,…): 580 mm
Depth inside (incubators,…): 330 mm
Lowest temperature (cooling): – 5 °C
BINDER Cooling incubator KB 53-230V KB 53UL-120V
Performance Data Temperature
Temperature range [°C] -5..100 -5..100
Temperature variation at 37 °C [± K] 0.2 0.2
Temperature fluctuation at 37 °C [± K] 0.1 0.1
Heating-up time to 37 °C [min] 5 5
Recovery time after 30 seconds door open at 37 °C [min] 2 2
Electrical data
Rated Voltage [V] 230 120
Power frequency [Hz] 50 60
Nominal power [kW] 0.6 0.6
Unit fuse [A] 10 12.5
Phase (Nominal voltage) [ph] 1 1
Outer dimensions not incl. fittings and connections
Width net [mm] 635 635
Height net [mm] 835 835
Depth net [mm] 580 580
Wall clearance back [mm] 100 100
Wall clearance sidewise [mm] 100 160
Doors
Inner doors 1 1
Unit doors 1 1
Internal Dimensions
Width [mm] 400 400
Height [mm] 400 400
Depth [mm] 330 330
Measures
Interior volume [L] 53 53
Net weight of the unit (empty) [kg] 78 78
permitted load [kg] 40 40
Load per rack [kg] 15 15
Environment-specific data
Energy consumption at 37 °C [Wh/h] 70 70
Sound-pressure level [dB(A)] 49 49
Fixtures
Number of shelves (std./max.) 2/4 2/4

All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.